Alpha-fetoprotein tumor marker, serum

Alphabetical Test listing

Alpha-fetoprotein tumor marker, serum-20

  
Alpha-fetoprotein tumor marker, serum
  
20
  
82105.0
  
AFT
  
AFP
Tumor marker
Alpha fetoprotein
  

The test method changed on 1/3/2023. If this test has been used for serial monitoring, rebaselining is recommended. Rebaselining consists of 2 measurements, collected 3-6 weeks apart.
The Roche Elecsys CEA assay is an electrochemiluminescence immunoassay "ECLIA" performed on the Roche Cobas "e" immunoassy analyzers.
Values obtained with different assay methods may be different and cannot be used interchangeably.

  

Not for pregnant females

Biotin supplements may cause clinically significant interference for this test assay. If interference is suspected, it is strongly recomended that biotin is discontinued for at least one week prior to retesting.

  
Serum
  
1.0 mL
  
0.5 mL
  

Immediatley following collection, mix sample by inverting 5 times

  
  1. Allow sample to clot for a minimum of 30 minutes
  2. Spin within two (2) hours of sample collection

Red:

  1. Allow sample to clot
  2. Spin
  3. Transfer serum to a False bottom plasma/serum transport vial/tube (AHL), labelled as serum, within two (2) hours of sample collection
  

Gold serum separator (SST) tube

False bottom plasma/serum transport vial/tube (AHL)

  
Lithium heparin (Li hep) plasma
  
  

Immediately follow collection, mix sample by inverting 8 - 10 times to prevent clotting

  

Spin within two (2) hours of sample collection

 

  

Lt green plasma separator (PST)

 

  

Refrigerated (preferred) - 14 days

Ambient - 5 days

Frozen - 6 months

  
  • Improper labels (unlabeled or mislabeled)
  • Hemolysis (some procedures)
  • Improper anticoagulant or ratio
  • Delay in transport
  • Improper storage temperature affecting results
  • Inappropriate timing of collection
  • Improper container
  • Leaking container resulting in compromised specimen
  • Quantity not sufficient (QNS)
  
AHL - Chemistry: C
  
Mo - Sa
  
2 days
  

Electrochemiluminescence immunoassay (ECLIA)

  

≤ 8.3 ng/mL

  
82105
  
Yes
  
  
11/30/2001
  
12/23/2022
  
11/09/2021