Cortisol, post Dexamethasone suppression test-68

Test info

  
Cortisol, post Dexamethasone suppression test
  
68
  
82533.1
  
DEX
  
Dexamethasone suppression
  

Biotin supplements may cause clinically significant interference for this test assay. If interference is suspected, it is strongly recommended that biotin is discontinued for at least one week prior to retesting.

Specimen

  
Serum
  
  
1.0 mL
  
0.5 mL
  
  • A number of protocols are available
    • The most common protocol is to administer 1.0 mg of Dexamethasone at 11 PM and to draw the specimen the next morning at 8 AM.
  • It is not recommeded to do this test on patients treated with Prednisolone.
  • Immediately following collection, thoroughly mix sample by gently inverting 5 times
  
  1. Allow sample to clot for a minimum of 30 minutes
  2. Spin within two (2) hours of sample collection
  

Gold serum separator (SST) tube

  
Lithium heparin (Li hep) plasma
  
  

Immediately following collection, mix sample thoroughly by gently inverting 8 - 10, times to prevent clotting

  

Spin within two (2) hours of sample collection

 

  

Plasma separator (Lt green PST)

 

  

Refrigerated (preferred) - 4 days

Ambient - 24 hours

Frozen - 12 months

  
  • Improper labels (unlabeled or mislabeled)
  • Hemolysis (some procedures)
  • Improper anticoagulant or ratio
  • Delay in transport
  • Improper storage temperature affecting results
  • Inappropriate timing of collection
  • Improper container
  • Leaking container resulting in compromised specimen
  • Quantity not sufficient (QNS)

Performance

  
AHL - Chemistry: C
  
Mo - Fr
  
1 - 2 days
  

Electrochemiluminescence immunoassay (ECLIA)

Clinical and Interpretive info

  

< 5.0 µg/dL

Billing

  
82533

Tracking

  
08/24/2004
  
12/26/2022
  
11/09/2021