Meningitis/Encephalitis (ME) pathogen multiplex PCR, CSF

Alphabetical Test listing

Meningitis/Encephalitis (ME) pathogen multiplex PCR, CSF-14504

  
Meningitis/Encephalitis (ME) pathogen multiplex PCR, CSF
  
14504
  
LAB14504
  
ME-CSF
  
ME, CSF panel
  
Bacteria:
Escherichia coli K1
Haemophilus influenza
Listeria monocytogenes
Neisseria meningitidis
Streptococcus agalactiae
Streptococcus pneumonia

Virus:
Cytomegalovirus
Enterovirus
Herpes simplex virus 1/Herpes simplex virus 2 (HSV1/2)
Human herpesvirus 6 (HHV-6)
Human parechovirus
Varicella zoster virus (VZV)


Fungal:
Cryptococcus neoformans/gattii

  

Rapid detection of meningitis and encephalitis caused by:

  • Escherichia coli K1 (K1 serotype only)
  • Haemophilus influenzae
  • Listeria monocytogenes
  • Neisseria meningitidis (encapsulated strains only)
  • Streptococcus agalactiae (Group B Strep)
  • Streptococcus pneumoniae
  • Cytomegalovirus (CMV)
  • Enterovirus
  • Herpes simplex virus 1 (HSV-1)
  • Herpes simplex virus 2 (HSV-2)
  • Herpes simplex virus 6 (HHV-6)
  • Human parechovirus
  • Varicella zoster virus (VZV)
  • Cryptococcus neoformans/gattii
  
  • Possible interference of assay performance, resulting in false-negative test results in patients on prior treatments. Test performance is dependent on appropriate sample collection, handling, transport, and storage. False-negative results may arise from genetic rearrangements or closely related strains of the target organisms in the panel.
  • This test is not indicative for Ventricular shunt infections
  • Not for a test of cure
  • This test cannot differentiate between Cryptococcus neoformans and Cryptococcus gattii
  • This test does not replace conventional CSF Gram stain and culture
    • Always Order Gram stain and culture (6557) when submitting this test
  
Cerebrospinal fluid (CSF)
  

Sterile screw top vial/container

  
0.4 mL
  
0.2 mL
  

CSF specimens should be collected via lumbar puncture

  

Do not be centrifuge, dilute, or otherwise process prior to testing

  

Sterile screw top vial/container

  

Refrigerated (preferred) - 7 days

Ambient - 24 hours

Frozen (≤ -15 °C)  - 30 days

  
1 week
  
  • Centrifuged specimen
  • Specimen not aliquoted aseptically
  • CSF collected via indwelling medical device (shunt)
  
AHL - Microbiology: M
  
Daily
  
4 hours from receipt at the performing laboratory
  

Multiplex Polymerase Chain Reaction (PCR)

  

Not detected (Negative)

  

Detected (positive) or invalid results for any targets

  
87483
  
06/20/2022
  
12/09/2022
  
12/18/2023