Almond (F20), IgE

Alphabetical Test listing

Almond (F20), IgE-4223

  
Almond (F20), IgE
  
4223
  
86003.1
  
F20
  
Allergen ImmunoCap
A31
  

As an aid in the clinical diagnosis of IgE mediated allergic disorders in conjunction with other clinical findings

  
Serum
  
  
1.0 mL
  
0.3 mL
  

For more than one allergen, use this formula to calculate the amount of serum required:

50 µL X (# of allergens) + 250 µL = total amount serum needed 

Immediatley following collection, mix sample by gently inverting 5 times

  
  1. Allow sample to clot for a minimum of 30 minutes
  2. Spin
  

Gold serum separator (SST) tube

  

Red serum vial/tube - 5 mL

Red/grey serum separator (SST) vial/tube, 10 mL

  
  1. Allow sample to clot
  2. Spin
  3. Transfer serum to a False bottom plasma/serum transport vial/tube (AHL) labelled as "serum"

 

  

False bottom screw cap serum/plasma transfer tube/vial (AHL)

Red/grey serum separator (SST) vial/tube, 10 mL

  

 Temperature

 Time

 Special container

 Ambient (preferred)

 3 days

 

 Refrigerated

 7 days

 

 Frozen

 OK, one time

False bottom plasma/serum transport vial/tube + cap (AHL)

  
  • Improper labels (unlabeled or mislabeled)
  • Hemolysis (some procedures)
  • Improper anticoagulant or ratio
  • Delay in transport
  • Improper storage temperature affecting results
  • Improper container
  • Leaking container resulting in compromised specimen
  • Quantity not sufficient (QNS)
  
AHL - Immunology: I-L
  
Mo - Fr
  
3 - 5 days
  

Fluorescence Enzyme Immunoassay (FEIA)

  

< 0.35 kU/L = Negative

  

Allergy class severity values

ImmunoCAP IGE Class Reference Values:

Traditionally, allergen tests have been reported with an associated class value of zero to 6, with Class 0 indicating an undetectable allergen level, and Class 6 indicating a strong positive allergen level.

The following chart correlates ImmunoCAP IgE allergen results in kU/L to the associated class, with a suggested interpretation. This chart is applicable to all allergen-specific tests at Allina Health Laboratory, which are performed exclusively on ImmunoCAP assays. Note that the Class Reference Values provided are specific for ImmunoCAP assays and may not be used to interpret allergy tests performed by other (non-ImmunoCAP) methods.

 

 Class  IgE kU/L  Interpretation
0  <0.35  Undetectable
1  0.35 - 0.70  Equivocal
2  0.71 - 3.50  Positive
3  3.51 - 17.50  Positive
4  17.51 - 50.00  Strong positive
5  50.01 - 100.00  Strong positive
6  >100.00  Strong positive
  
86003
  
Yes
  
  
06/30/2008
  
01/31/2022
  
11/12/2021